PROVA 123 PROCESS CALIBRATOR
• 4-20mA (1KOhm load, 24V loop supply)
• 0-100.00mV, 0-1.0000V, 0-12.000V
• K, J, E, T Thermocouples (°C and °F)
• Frequency 1-62500Hz
• 0.025% Basic Accuracy
• Easy Key-pad Operation
• Easy Step and Auto Ramp Functions
• 0-20mA, 0-24mA selectable
• 1-100% Input (mA, mV, V)
• Beeper Warning when Output is Open (mA)
Electrical Specification (23°C ± 5°C, 3 minutes after power is on)
1.mA DC Current (1K Max. Load, 24V Loop Supply)
Range: 4 - 20mA, 0 - 20mA, 0 -24 mA
Resolution : 1 µA
Accuracy : ± 0.025% ±3 µA
"Beeper" warning when output is open and specified current output > 1mA
2.mV, V DC Voltage (1mA Supply Current)
Range: 0-100.00mV - 0-1.0000V
Resolution: 10µV- 100µV
Accuracy: ± 0.025% ± 3µA
"Beeper" warning when output is short and specified voltage output > 10mV
3.K, J, E, T type Thermocouples( 1 C,1 F Resolution,1K load Min. )
Thermocouple Range /Accuracy:
K : -200 to 0ºC ±1.1ºC ( -328 to 32ºF ± 2.0ºF)
K : -0 to 1370ºC ±0.8ºC ( 32 to 2400ºF ± 1.5ºF)
J : -100 to 0ºC ±0.9ºC ( -148 to 32ºF ± 1.6ºF)
J : -0 to 760ºC ±0.7ºC ( 32 to 1400ºF ± 1.2ºF)
E : -100 to 0C ±0.9ºC( -148 to 32ºF ± 1.6ºF)
E : 0 to 700ºC ±0.7ºC ( 32 to 1292ºF ± 1.2ºF)
T : -200 to 0ºC ±1.0ºC( -328 to 32ºF ± 1.8ºF)
T : 0 to 400ºC ±0.8ºC ( 32 to 752ºF ± 1.5ºF)
4.Frequency (1 - 125Hz, 1K ohm Load Min.)
Range : 1 - 125 Hz
Resolution : 1 Hz
Accuracy : ± 0.04H
Available Frequencies(126 - 62500Hz, Accuracy ± 0.01% ± 0.04Hz, 1K Load Min)
General Specifications:
Battery Type: 9V Alkaline battery
Display: 4 x 5 Digits
Operating Temperature: 0 to 50°C (32 to 122°F)
Operating Humidity: less than 85% relative
Storage Temperature: -20 to 60°C (-4 to 140°F)
Storage Humidity: less than 85% relative
Dimension: 88 x 168 x 26mm (3.46” x 6.61” x 1.03”)
Weight: 330g / 11.63 oz.